Thermo Scientific 840-297000
-
Description
-
Features
-
Specifications
<
>
The GENESYS 50 is a dual beam UV-Vis spectrophotometer in a single-cell configuration that meets all of the specifications for critical quality control applications and precise chemical measurements. The GENESYS 40 Dual Beam Vis Spectrophotometer is a visible spectral region version in this instrumental class. Both feature data resolution from 0.2 nm to 5 nm.
- 7-inch high-resolution touchscreen user interface
- Local control for speed and reliability
- Networking capable (Wi-Fi adaptor available)
- Large sample compartment with front access
Baseline Flatness |
±0.005A |
Compartments |
Accessible from top, front or side Able to accommodate cells up to 100mm cells Available accessory for test tubes up to 25 mm diameter and 150 mm height Removable, washable sample compartment liner with magnetic placement and hold-down |
Connections |
Single USB-A supports flash memory devices for method and data storage Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, mouse Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available Print via USB, Ethernet or Wi-Fi USB adaptor available |
Description |
GENESYS 40 Visible Spectrophotometer |
Detector Type |
Dual Silicon Photodiodes |
Dimensions (L x W x H) |
35.5 x 38.5 x 19.5 cm |
Display |
7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
Drift |
(At 500 nm after 1 hour warmup) <0.0010A/Hr |
Electrical Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100–240 volts, 50–60 Hz. |
Keypad |
Touchscreen |
Lamp |
Tungsten-Halogen (>1000 hours) |
Noise |
(RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Optical Design |
Dual beam |
Photometric Accuracy Instrument |
±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A |
Photometric Display |
-3A to +5A |
Photometric Range |
–3A to +3.5A |
Photometric Repeatability |
±0.001A at 1A measured at 1.0A at 546 nm |
Printer |
Snap-on printer available |
Spectral Bandwidth |
5 nm |
Stray Light |
<0.05%T at 340 nm and 400 nm |
Wavelength Accuracy |
±0.5 nm |
Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range |
325 nm – 1100 nm |
Wavelength Repeatability |
< ±0.2 nm |
Wavelength Scan Speed |
Automatic – up to 1800 nm/min |
Weight (Metric) |
7.5 kg |
Warranty |
2 years standard manufacturer's with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase |