High Performance Sugar Analysis, Now Easier Than Ever. It has become common that dark color - raw sugar samples are measured at 882nm without decolorization treatment.

SAC-i 589/882 can measure samples at both 589nm and 882nm by changing the optical units

Add to Cart

ATAGO 5952 Specifications

ScaleAngle of Rotation, International Sugar Scale (without temperature compensation), International Sugar Scale (with Automatic Temperature Compensation)
Measurement ReadingsAngle of Rotation, International Sugar Scale (without temperature compensation), International Sugar Scale (with Automatic Temperature Compensation), Specific Rotation, Concentration, Purity, User scale, Angle of Rotation (Temperature Compensation of Quartz Plate), Temperature (°F/°C)
Measurement RangeAngle of Rotation -89.9999 to +90° or -360 to +360°, International Sugar Scale -259 to +259°
ResolutionAngle of Rotation: 0.0001°, International Sugar Scale: 0.0001°Z
Measurement Accuracy (nD)589nm
Angle of Rotation: ±0.005° (-45 to +45°)
International Sugar Scale: ±0.015°Z (-130 to +130°Z)
Other ranges
Angle of Rotation: ±0.01°
International Sugar Scale: ±0.03°Z
* Checked by reading a standard quartz plate
882nm
Angle of Rotation: ±0.005° (-19.3 to +19.3°)
International Sugar Scale: ±0.015°Z (-130 to +130°Z)
Other ranges
Angle of Rotation: ±0.01°
International Sugar Scale: ±0.07°Z
* Checked by reading a standard quartz plate
Temperature Compensation Range50 to 104°F (10 to 40°C)
Ambient TemperatureUse the instrument where the temperature is between 50 to 104°F (10 to 40°C)
Display Panel7.5in (19.05cm) color LCD (touch screen)
±0.2°C
Light SourceLED: 589nm interference filter, 882nm interference filter
PrismGlan-Taylor Prism
Measurement Wavelength589nm (equivalent to the sodium-D line spectrum), 882nm (infrared nearby)
Power SupplyAC100 to 240V, 50/60Hz
Dimensions & Weight23.62 x 14.37 x 8.27in (60 x 36.5 x 21cm), 44.09lbs (20kg)
Power consumption/td>140VA