- Store
- >
- Refurbished & Used
- >
- Test & Measurements
- >
- TESTO
- >
- Testo 340 - Combustion Analyzer for Commercial and Industrial Applications, Used
Testo 340 - Combustion Analyzer for Commercial and Industrial Applications, Used
SKU:
Testo 340
$4,550.00
$4,250.00
$4,250.00
Unavailable
per item
Sold As Is. Boots up, All testing functions work properly. ” . Used in working Condition
The Testo 340 is the ideal tool to confirm proper set-up and to identify emission problems before they get serious. Oxygen is the base sensor in all Testo 340 Analyzers. Simply add one, or up to three sensors for your specific job.
Sold out
Testo 340 Flue Gas Analyzer
This combustion analyzer can be configured in many ways for commissioning commercial boilers, testing engines in midstream gas pipelines, tuning industrial burners in refineries, or cement or glass or any other combustion system that requires high confidence, so you can make the proper control adjustments.
- The back-lit display is easily customized, can display from 4 to 8 lines, or show a variety of measurement options, such as PPM values or as MMBtu/hr. shows measurements
- Important calculation including, combustion efficiency, excess air calculations, corrected NOx and corrected CO values.
- 18 fuels to choose from
- On board Diagnostics – Simple Push the I button and scroll down to see instrument status.
- Draft and differential pressure and Mass Emission calculations (i.e. lbs./hr., tons/yr) with any Pitot tube
Differential Pressure
Measuring Range | -80 to 80 InH₂O / -200 to 200 hPa |
Accuracy | ±0.2 InH₂O (-20.0 to 20.0 InH₂O) / ±0.5 hPa (-49.9 to 49.9 hPa) ±1.5 % of mv (Remaining Range) |
Resolution | 0.0 InH₂O / 0.1 hPa |
Absolute Pressure
Measuring Range | +240 to +460 InH₂O / +600 to +1150 hPa |
Accuracy | ±4 InH₂O / ±10 hPa |
Resolution | 1 InH₂O / 1 hPa |
Flue Gas O₂
Measuring Range | 0 to 25 Vol.% |
Accuracy | ±0.2 Vol.% |
Resolution | 0.01 Vol.% |
Reaction Time t₉₀ | < 20 sec |
Flue Gas CO
Measuring Range | 0 to 10000 ppm |
Accuracy | ±10 ppm or ±10 % of mv (0 to 200 ppm) ±20 ppm or ±5 % of mv (201 to 2000 ppm) ±10 % of mv (2001 to 10000 ppm) |
Resolution | 1 ppm |
Reaction Time t₉₀ | < 40 sec |
Flue Gas COlow
Measuring Range | 0 to 500 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of mv (Remaining Range) data corresponds to 20°C ambient temperature. Additional temperature coefficient 0.25% of reading/K. |
Resolution | 0.1 ppm |
Reaction Time t₉₀ | < 40 sec |
Flue Gas NO
Measuring Range | 0 to 4000 ppm |
Accuracy | ±5 ppm (0 to 99 ppm) ±5 % of mv (100 to 1999 ppm) ±10 % of mv (2000 to 4000 ppm) |
Resolution | 1 ppm |
Reaction Time t₉₀ | < 30 sec |
Flue Gas NOlow
Measuring Range | 0 to 300 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of mv (Remaining Range) |
Resolution | 0.1 ppm |
Reaction Time t₉₀ | < 30 sec |
Flue Gas NO₂
Measuring Range | 0 to 500 ppm |
Accuracy | ±10 ppm (0 to 199 ppm) ±5 % of mv (Remaining Range) |
Resolution | 0.1 ppm |
Reaction Time t₉₀ | < 40 sec |
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.
Flue Gas SO₂
Measuring Range | 0 to 5000 ppm |
Accuracy | ±10 ppm (0 to 99 ppm) ±10 % of mv (Remaining Range) |
Resolution | 1 ppm |
Reaction Time t₉₀ | < 40 sec |
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.
Flue Gas Draft
Measuring Range | -16 to +16 InH₂O / -40 to +40 hPa |
Accuracy | ±0.01 InH₂O (-1.20 to +1.20 InH₂O) / ±0.03 hPa (-2.99 to +2.99 hPa) ±1.5 % of mv (Remaining Range) |
Resolution | 0.01 InH₂O / 0.01 hPa |
Temperature
Measuring Range | -40.0° to 2192.0 °F / -40 to +1200 °C |
Accuracy | ±0.9 °F (32.0° to 210.2 °F) / ±0.5 °C (0 to +99 °C) ±0.5 % of mv (Remaining Range) |
Resolution | 0.1 °F / 0.1 °C |
Efficiency (calculated)
Measuring Range | 0 to 120 % |
Resolution | 0.1 % |
Flue Gas Loss (calculated)
Measuring Range | 0 to 99.9 % |
Resolution | 0.1 % |
Flue Gas Dewpoint (calculated)
Measuring Range | 0.0° to 99.9 °Ctd |
Resolution | 0.1 °Ctd |
Flue Gas CO₂ Calculation (calculated from O₂)
Measuring Range | 0 to CO₂ max |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Reaction Time t₉₀ | < 40 sec |
General Technical Data
Weight | 34 oz. / 960 g |
Dimensions | 11 x 4 x 3 in. / 283 x 103 x 65 mm |
Operating Temperature | 23° to 122 °F / -5 to +50 °C |
Housing | TPE PC |
Protection Class | IP40 |
Product Colour | Black |
Connectable Probes | 1 x flue gas probe; 1 x temperature probe; 1 x differential pressure |
Battery Life | > 6h |
Display Type | Graphic Display |
Display Size | 160 x 240 pixels |
Power Supply | Battery block 3.7 V / 2.4 Ah, Mains unit 6.3 V / 2 A |
Maximum Negative Pressure/Flue Gas | -2.901 psi / -200 mbar |
Maximum Positive Pressure/Flue Gas | 0.725 psi / 50 mbar |
User Defined Fuels | 10 user-defined fuels incl. test gas as fuel |
House Length | max. 7.8 m (corresponds to two probe hose extensions) |
Pump Flow | 0.6 l/min (regulated) |
Maximum | 100 folders |
Storage per Folder | Max. 10 sites |
Storage per Site | Max. 200 logs |
Miscellaneous | The max. number of logs is determined by the number of folders or sites |
Interface | Bluetooth®; USB; IR/IRDA interface; gas outlet; Mains connection; probe input/ probe inputs; Differential Pressure |
Storage Temperature | -4° to 122 °F / -20 to +50 °C |