Scinteck Instruments
  • Store
  • About Us
  • Contact Us
  • Downloads
  • Store
  • >
  • New - Buffer
  • >
  • Refurbished & Used
  • >
  • Test & Measurements
  • >
  • TESTO
  • >
  • Testo 340 - Combustion Analyzer for Commercial and Industrial Applications, Used

Testo 340 - Combustion Analyzer for Commercial and Industrial Applications, Used

SKU: Testo 340
$4,550.00
$4,250.00
$4,250.00
Unavailable
per item

Sold As Is. Boots up, All testing functions work properly. ” . Used in working Condition


The Testo 340 is the ideal tool to confirm proper set-up and to identify emission problems before they get serious. Oxygen is the base sensor in all Testo 340 Analyzers. Simply add one, or up to three sensors for your specific job.

Sold out
Add to Cart
Testo 340 Flue Gas Analyzer

The Testo 340 is the ideal tool to confirm proper set-up and to identify emission problems before they get serious.  Oxygen is the base sensor in all Testo 340 Analyzers.  Simply add one, or up to three sensors for your specific job.
 
This combustion analyzer can be configured in many ways for commissioning commercial boilers, testing engines in midstream gas pipelines, tuning industrial burners in refineries, or cement or glass or any other combustion system that requires high confidence, so you can make the proper control adjustments.

  • The back-lit display is easily customized, can display from 4 to 8 lines, or show a variety of measurement options, such as PPM values or as MMBtu/hr. shows measurements
  • Important calculation including, combustion efficiency, excess air calculations, corrected NOx and corrected CO values.  
  • 18 fuels to choose from
  • On board Diagnostics – Simple Push the I button and scroll down to see instrument status.
  • Draft and differential pressure and Mass Emission calculations (i.e. lbs./hr., tons/yr) with any Pitot tube

Differential Pressure

Measuring Range
-80 to 80 InH₂O / -200 to 200 hPa
Accuracy
±0.2 InH₂O (-20.0 to 20.0 InH₂O) / ±0.5 hPa (-49.9 to 49.9 hPa)
±1.5 % of mv (Remaining Range)
Resolution
0.0 InH₂O / 0.1 hPa

Absolute Pressure

Measuring Range
+240 to +460 InH₂O / +600 to +1150 hPa
Accuracy
±4 InH₂O / ±10 hPa
Resolution
1 InH₂O / 1 hPa

Flue Gas O₂

Measuring Range
0 to 25 Vol.%
Accuracy
±0.2 Vol.%
Resolution
0.01 Vol.%
Reaction Time t₉₀
< 20 sec

Flue Gas CO

Measuring Range
0 to 10000 ppm
Accuracy
±10 ppm or ±10 % of mv (0 to 200 ppm)
±20 ppm or ±5 % of mv (201 to 2000 ppm)
±10 % of mv (2001 to 10000 ppm)
Resolution
1 ppm
​Reaction Time t₉₀
< 40 sec

Flue Gas COlow

Measuring Range
0 to 500 ppm
Accuracy
±2 ppm (0 to 39.9 ppm)
±5 % of mv (Remaining Range) data corresponds to 20°C ambient temperature. Additional temperature coefficient 0.25% of reading/K.
Resolution
0.1 ppm
Reaction Time t₉₀
< 40 sec

Flue Gas NO

Measuring Range
0 to 4000 ppm
Accuracy
±5 ppm (0 to 99 ppm)
±5 % of mv (100 to 1999 ppm)
±10 % of mv (2000 to 4000 ppm)
Resolution
1 ppm
Reaction Time t₉₀
< 30 sec

Flue Gas NOlow

Measuring Range
0 to 300 ppm
Accuracy
±2 ppm (0 to 39.9 ppm)
±5 % of mv (Remaining Range)
Resolution
0.1 ppm
Reaction Time t₉₀
< 30 sec

Flue Gas NO₂

Measuring Range
0 to 500 ppm
Accuracy
±10 ppm (0 to 199 ppm)
±5 % of mv (Remaining Range)
Resolution
0.1 ppm
Reaction Time t₉₀
< 40 sec
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.

Flue Gas SO₂

Measuring Range
0 to 5000 ppm
Accuracy
±10 ppm (0 to 99 ppm)
±10 % of mv (Remaining Range)
Resolution
1 ppm
Reaction Time t₉₀
< 40 sec
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.

Flue Gas Draft

Measuring Range
-16 to +16 InH₂O / -40 to +40 hPa
Accuracy
±0.01 InH₂O (-1.20 to +1.20 InH₂O) / ±0.03 hPa (-2.99 to +2.99 hPa)
±1.5 % of mv (Remaining Range)
Resolution
0.01 InH₂O / 0.01 hPa

Temperature

Measuring Range
-40.0° to 2192.0 °F / -40 to +1200 °C
Accuracy
±0.9 °F (32.0° to 210.2 °F) / ±0.5 °C (0 to +99 °C)
±0.5 % of mv (Remaining Range)
Resolution
0.1 °F / 0.1 °C

Efficiency (calculated)

Measuring Range
0 to 120 %
Resolution
0.1 %

Flue Gas Loss (calculated)

​Measuring Range
0 to 99.9 %
Resolution
0.1 %

Flue Gas Dewpoint (calculated)

Measuring Range
0.0° to 99.9 °Ctd
Resolution
0.1 °Ctd

Flue Gas CO₂ Calculation (calculated from ​O₂)

​Measuring Range
0 to CO₂ max
Accuracy
±0.2 Vol.%
Resolution
0.1 Vol.%
Reaction Time t₉₀
< 40 sec

General Technical Data

Weight
34 oz. / 960 g
Dimensions
11 x 4 x 3 in. / 283 x 103 x 65 mm
Operating Temperature
23° to 122 °F / -5 to +50 °C
Housing
TPE PC
Protection Class
IP40
Product Colour
Black
Connectable Probes
1 x flue gas probe; 1 x temperature probe; 1 x differential pressure
Battery Life
> 6h
Display Type
Graphic Display
Display Size
160 x 240 pixels
Power Supply
Battery block 3.7 V / 2.4 Ah, Mains unit 6.3 V / 2 A
Maximum Negative Pressure/Flue Gas
-2.901 psi / -200 mbar
Maximum Positive Pressure/Flue Gas
0.725 psi / 50 mbar
User Defined Fuels
10 user-defined fuels incl. test gas as fuel
House Length
max. 7.8 m (corresponds to two probe hose extensions)
Pump Flow
0.6 l/min (regulated)
Maximum
100 folders
Storage per Folder
Max. 10 sites
Storage per Site
Max. 200 logs
Miscellaneous
The max. number of logs is determined by the number of folders or sites
Interface
Bluetooth®; USB; IR/IRDA interface; gas outlet; Mains connection; probe input/ probe inputs; Differential Pressure
Storage Temperature
-4° to 122 °F / -20 to +50 °C


Company

About Us


Support

Contact Us
Terms of Use
© 2021 Scinteck Instruments, LLC. All Rights Reserved
  • Store
  • About Us
  • Contact Us
  • Downloads