Waters LCT Premier XE MS with Water UPLC System
ONLY AVAILABLE IN PAKISTAN CONTACT SCINTECK TECHNOLOGY KARACHI PAKISTAN AT [email protected] OR PHONE 0301-0185115
Includes:
- Waters LCT Premier XE TOF-MS
- ESI Source
- Masslynx Data System
- TUV Detector
- Binary Solvent manager
- Sample manager
- Software Loaded PC, Mouse, Keyboard and Monitor included
Sold out
Waters LCT Premier XE MS The Waters LCT Premier XE orthogonal acceleration time-of-flight (oa-TOF), benchtop is a market leading mass spectrometer for LC Applications.
The LCT Premier XE's high sensitivity Z-Spray ion source, ion transfer optics and TOF analyzer provides the most sensitive benchtop oa-TOF instrument in its class, giving capability of detecting very low level components with full spectral information. Enhanced resolution (up to 15,000 FWHM over 30,000 m/z) spectral data provides the capability of high accuracy exact mass measurements providing improved data specificity through elemental composition calculations resulting in an increased confidence in your data. For challenging and complex samples, the rapid spectral acquisition rate and wide dynamic range provide optimized detection capabilities making it the ideal MS instrument for qualitative applications.
The Waters LCT Premier XE provides a powerful platform for MS analyses, upon which total system solutions may be designed to maximize data quality for your application. Its small footprint sits easily on your laboratory benchtop and is optimized for simplified fluidic connections to your LC system. When combined with ACQUITY UPLC, the LCT Premier XE provides optimized full spectral MS detection for UPLC separations, providing unsurpassed speed sensitivity and resolution for high throughput applications
Features:
- Resolution: up to 15,000 FWHM over 30,000 m/z
- Mass Accuracy: < 3 mDa
- High sensitivity Z-Spray ion source
- Wide dynamic range
The Waters ACQUITY UPLC System provides an integrated configuration for solvent and XYZZ sample management designed for use with ACQUITY UPLC chemistries. The core ACQUITY UPLC System comprises a Binary Solvent Manager, a Sample Manager and a Solvents Tray.
Includes:
- TUV Detector or PDA Detector ($1000 more)
- Sample Manager
- Binary Solvent Manager
- 15000 psi Max Pressure
- 1034 bars Max Pressure
- Computer Loaded with Empower
Waters Acquity TUV Detector UPLC TUV DetectorThe ACQUITY UPLC TUV Detector is a tunable, dual wavelength ultraviolet/visible (UV/Vis) detector that offers optimal linearity, resolution, and sensitivity for UPLC/UV separations. It features uniquely designed, patented, light-guiding flow cell technology, low-noise performance (6 μAU), and support for data rates up to 80 Hz. The ACQUITY UPLC TUV Detector, coupled with our innovative ACQUITY UPLC Column chemistries and separations modules, leverages ultra-low dwell volume and low-noise, high-speed detection to bring productivity, sensitivity, and resolution that are unprecedented in HPLC techniques being used in the laboratory today.
Specifications:
- Wavelength range: 190 to 700 nm
- Wavelength accuracy: ±1 nm
- Linearity range: Deviation at 2.5 AU ≤5%, propylparaben, at 257 nm
- Baseline noise:6 x 10^-3 AU, at 230 nm, 10 mm shunt cell, 1.0 second filter time constant
- Drift: ≤5.0 x 10^-4 AU/hour/°C at 230 nm
- Sampling rate: Up to 80 points/s
- Unattended operation: Leak sensors, full diagnostic data captured through console software
- Light source: Prealigned, intelligent technology Deuterium lamp
- Flow cell design: Light-guiding UPLC flow cell with intelligent chip technology
- Path length: 10 mm (analytical flow cell)
- Cell volume: 500 nL (analytical flow cell)
- Pressure limit: 1000 psi (all flow cells)
- Wetted materials: 316 stainless steel, fused silica, Teflon AF, PEEK (analytical flow cell)
- Optical bandwidth: 5 nm
Waters Acquity Binary Solvent Manager UPLC Binary Solvent ManagerSpecifications:
- Number of solvents: Up to four: in any combination of two: Al or A2 and Bl or B2
- Solvent storage: Solvents Tray accommodates up to four chromatographic solvents, two Sample Manager wash solvents and one Binary Solvent Manager seal wash solvent
- Solvent conditioning: Vacuum degassing (Six-channel): one channel per solvent, and two channels for Sample Manager wash solvents
- Operating flow rate range: 010 to 2.000 mL/min, in 0.001 mL increments
- Compressibility compensation: Automatic and continuous
- Effective system delay volume: <120 ulL, independent of system backpressure (with standard 50-ilL mixer installed)
- Plunger seal wash: Integral, active, programmable
- Gradient profiles: Eleven gradient curves [including linear, step (2), concave (4), and convex (4)]
- Wet prime: Automatic
- Maximum operating pressure: 15,000 psi up to 1 mL/min, 9,000 psi up to 2 mL/min per pump, not more than 15,000 psi total
- Composition accuracy:
- ±0.5% absolute (full scale) from 5 to 95% of flow rates
- from 0.5 to 2.0 mLlmin (Contact Waters for conditions used)
- Composition precision: 0.15% RSO or ±0.04 min SO, whichever is greater, based on retention time
- Flow precision: 0.075% RSO or ±0.02 min SO, six replicates, based on retention time or volumetric measures (0.500 to 2.000 ml/min)
- Flow accuracy: ± 1.0% at 0.5 mL/min with degassed methanol, per Waters AQT/SystemsQT protocol
- Primary wetted materials: 316 stainless steel, UHMWPE, sapphire, ruby, FEP, PTFE, ETFE, diamond-like coating, PEEK and PEEK alloys, titanium alloys
- Unattended operation: Leak sensors, full diagnostic data captured through console software
Waters Acquity Sample manager UPLC Sample ManagerSpecifications:
- Number of sample plates: Total of two plates, expandable to up to 22 plates with optional Sample Organizer (see below):
- 96 and 384 microtiter plates
- 48 position 2.00-mL vial plates
- 48 position 0.65-mL micro-centrifuge tube plates
- 24 position 1.50-mL micro-centrifuge tube plates
- Maximum sample capacity: 768 in two 384-well plates; expandable to up to 8,448 samples with optional Sample Organizer
- Number of sample injections: 1 to 99 injections per sample
- Injection volume range: 0.1 to 50.0 iJL, in 0.1 iJL increments, partial or full loop mode; 10ul loop is standard; 1, 2, 5, 20, and 50-uL loops also available
- Sample delivery precision(full loop injection mode): < 0.3% RSO, full loop, standard lO-iJL loop (default wash/purge conditions), per Waters AQT/SystemsQT protocol
- Sample delivery precision(PLNO injection mode): < 1% RSO within 20% to 75% of loop volume for 1, 2, 5, 10, 20, and 50-iJL loops, UV detection
- Injector linearity: > 0.999 coefficient of deviation (from 20% to 75%, partial loop overfill mode (PLNO injection mode), per Waters' AQT/SystemsQT protocol
- Sample temperature control:0 to 40.0 ·C, settable in 0.1 ·C increments (assumes an ambient temperature of 25.0 ·C). At an ambient temperature of 21.0 ·C or lower the sample manager will maintain the temperature of the sample compartment down to 4.0 ·C with a tolerance of -2.0/+6.0 ·C, when configured with the maximum number of vials and/or plates
- Injection cycle time:
- < 15 sec between multipe injections with "load ahead" enabled
- 30 sec with single weak wash, lO-iJL loop, pressure assist mode
- Sample probe: XYZZ based needle-in-needle design
- Minimum sample required: 5 uL residual, using maximum recovery 2-mL vials (zero offset)
- Wash solvents: Two degassed: strong solvent and weak wash solvent, programmable to suit application
- Sample carryover: < 0.005% or < 2.000 nL, whichever is greater
- Advanced operations: Loop off-line mode, load ahead
- Unattended operation: Leak sensors, full diagnostic data control captured through console software
- Primary wetted material: Titanium alloy, 316 stainless steel, fluoropolymer, fluoroelastomer, PPS alloy, PEEK alloy, PPS, PEEK, OLC coating, gold