- Store
- >
- New
- >
- Spectroscopy
- >
- Persee
- >
- Persee T6 Series UV-Vis Spectrophotometer
Persee T6 Series UV-Vis Spectrophotometer
SKU:
$3,160.00
2340
3350
$2,340.00 - $3,350.00
Unavailable
per item
-
Overview
-
Key Features
-
Specifications
<
>
Designed for accuracy and performance in a compact design the T6 Series of
UV-Vis Spectrophotometers offer wavelength accuracy as low as +/- 1 nm combined
with low stray light of less than 0.05%T.
With a focus on reliability in a compact footprint, PERSEE’s entry level Spectrometers set a new bar for value and capability.
PERSEE delivers the capabilities required by customers in education, agricultural, environmental and food & beverage industries, all the options and services at such value that customers can get months of consumables or additional instrumentation within their budget without sacrificing performance.
UV-Vis Spectrophotometers offer wavelength accuracy as low as +/- 1 nm combined
with low stray light of less than 0.05%T.
With a focus on reliability in a compact footprint, PERSEE’s entry level Spectrometers set a new bar for value and capability.
PERSEE delivers the capabilities required by customers in education, agricultural, environmental and food & beverage industries, all the options and services at such value that customers can get months of consumables or additional instrumentation within their budget without sacrificing performance.
- High performance fixed 2nm spectral bandwidth
- Low stray light 0.05%T (T6U)
- Wavelength accuracy +/- 1nm (T6U)
- Supplied with either universal path-length 5 cell changer or fixed path length 8 cell changer as standard
- Holographic blazed grating 1200 lines/mm
- Local control software for photometric fixed wavelength measurement
- Easily upgraded to include quantitative analysis, multi wavelength spectrum & kinetics
- Built in cell holder storage
- Robust modular design with a small footprint
- Can be used with UV-WIN software (optional)
Optical system |
Split beam |
Scan speed |
Selectable |
Wavelength range |
190-1100 nm |
Wavelength accuracy |
T6U: ±1 nm T6V: ±2 nm |
Wavelength reproducibility |
T6U≤0.2 nm T6V≤0.4 nm |
Spectral bandwidth |
2 nm |
Photometric mode |
Transmittance, Absorbance, Energy |
Photometric range |
-0.3-3.0 Abs |
Photometric accuracy |
0.002 A (0-0.5 A), 0.004 A (0.5-1 A), 0.3 %T (0-100 %T) |
Photometric reproducibility |
00.001 A (0-0.5 A), 0.002 A (0.5-1 A), 0.15 %T (0-100 %T) |
Photometric noise |
0.001 A (500 nm, 30 min warm-up) |
Baseline flatness |
0.002 A (200-1000 nm) |
Baseline stability |
0.001 A/h (500 nm, 0 Abs, 2 hr warm-up) |
Stray light |
T6U: ≤0.05 %T (220 nm NaI, 340 nm NaNO2) T6V: ≤0.10 %T (220 nm NaI, 340 nm NaNO2) |
Standard functionality |
Photometric Measurement (Quantitative, Multi-wavelength, Spectrum and Kinetic measurements with program cards) |
Detector |
Silicon photo diode |
Light source |
Tungsten Halogen and Deuterium arc lamps |
Display |
Digital LCD display |
Printer |
Mini printer |
PC interface |
RS232 |
Software support |
Local and UV Win |
Power supply |
Switch mode 95-250 VAC 50-60 Hz |
Weight |
37 lbs |
Dimension |
22 x 20 x 13 in |