The AND Weighing MF series moisture analyzers are ideal for food, pharmaceutical, chemical processing, or any moisture analysis application, particularly those needing GLP/GMP/ISO compliance. This one-of-a-kind drying system, combined with an advanced heat protection mechanism for the weigh cell, provides superior accuracy and repeatability.
A&D® MF-50 - Moisture Balance
The AND Weighing MF series moisture analyzers are ideal for food, pharmaceutical, chemical processing, or any moisture analysis application, particularly those needing GLP/GMP/ISO compliance. This one of a kind drying system, combined with an advanced heat protection mechanism for the weigh cell, provides superior accuracy and repeatability.
Special Features:- Fast & uniform heating with halogen lamp and uniquely designed SRA (Secondary Radiation Assist) filter
- WinCT-Moisture Software (standard with MS & MX, option with MF) displays a real-time graph of moisture content rate change
- Sodium Tartrate Dihydrate is provided as a stable moisture content for accuracy checks of the analyzer
- Five Measurement Programs: Standard, Automatic, Timer, Manual, Quick
- Clear and easy-to-see VFD display
- Glass window for heating check and easy cleaning
- Self-check function
- Quick Reference Card
- Standard bi-directional RS-232C
- Conformity to GLP, GMP, GCP and ISO with time/date, ID, calibration data and check record outputs
- Full Digital Calibration
- Low Maintenance Cost
- 5 Year Manufacturer's Warranty
A&D - MF-50 | |
---|---|
Capacity | 51 g |
Readability | 0.002 g |
Pan Size | 85 mm dia. / 2.16" dia. |
Measurement Method | 400W straight halogen lamp heating system with SRA filter and SHS weighing technology |
Moisture Content Display | 0.05% / 0.1% / 1% |
Moisture Content Accuracy over 1g (Standard Deviation) | 0.20% |
Moisture Content Accuracy over 5g (Standard Deviation) | 0.05% |
Heating Technology | Halogen lamp (Straight type, 400 Watt max, 5000 hours) |
Drying Temperature (1°C increment) | 30°C - 200°C / 86°F - 392°F |
Memory of Measurement Programs | 10 sets |
Measurement Programs | Standard Mode/Automatic Mode/Quick Mode/Timer Mode/Manual Mode |
Measurement Mode | Moisture content(Wet or Dry base) /Dry content/Ratio/Weight |
Heating Mode | Standard / Quick / Step / Ramp |
Display Type | Large VFD |
Interface | RS-232C standard |
Data Memory Function | 50 |
Operating Temperature | 5°C - 40°C / 41°F - 104°F, less than 85%RH |
GLP / GMP / ISO | Available |
Self Check Function | Standard |
Communication Software | WinCT standard |
Power | AC 100V to 120V (3A) or AC 200V to 240V (1.5A), 50/60 Hz, Approx. 400W |
Scale Dimension (W x D x H) | 215 x 320 x 173 mm / 8.46 x 12.60 x 6.81" |
Weight | Approx. 6 kg / 13.2 lbs. |
Item No | Accessory | |
---|---|---|
AX-31 | Sample Pans (85 mm dia. washable x 100 pcs) | |
AX-32-2 | Glass Fiber sheet (78 mm dia. x 100 sheets) | |
AX-33 | Test Sample (Sodium Tartrate Dihydrate, 30 g x 12 pcs) | |
AX-34 | Halogen Lamp (replacement) 120V | |
AX-MX-34-240V | Halogen Lamp (replacement) 240V | |
AX-35 | Pan Handle (2 pcs) | |
AX-36 | Tweezers (2 pcs) | |
AX-37 | Spoon (2 pcs) | |
AX-38 | Display Cover (5 pcs) | |
AX-39 | Dust Cover | |
AX-42 | WinCT-Moisture (CD-ROM, Application Software for Windows) (standard for MS-70 & MX-50, option for MF-50) | |
AD-1700-20G | Calibration Mass (20 g) |